Medical Devices Regulation Art. 87 and IVDR Art. 82 where the texts differ

textually similar (0.965) · paired on the Medical Devices Regulation text of 19 July 2026 and the IVDR text of 10 January 2025 · headings the same

Medical Devices Regulation Art. 87 and IVDR Art. 82 are textually similar: their words match at a ratio of 0.965, where 1 is identical text. That says nothing about what either provision requires, or whether one should change because the other did.

Where the texts differ

Each passage is a differing span with up to 6 unchanged words either side, from the text as compared, spacing collapsed, and cut at 400 characters where it ends […]. It is an excerpt for reading where the two differ, not the verbatim text.

PassageMedical Devices Regulation Art. 87IVDR Art. 82
1

Article 87 Reporting of serious incidents and field

Article 82 Reporting of serious incidents and field

2

safety corrective actions 1. Manufacturers of devices made available on the Union market,

safety corrective actions 1. Manufacturers of devices, made available on the Union market,

3

on the Union market, other than investigational devices, shall report, to the relevant competent

on the Union market, other than devices for performance study, shall report, to the relevant competent

4

competent authorities, in accordance with Articles 92(5) and (7), the following: (a) any

competent authorities, in accordance with Articles 87(5) and (7), the following: (a) any

5

on the Union market, except expected side-effects which are clearly documented in the

on the Union market, except expected erroneous results which are clearly documented and quantified

6

expected side-effects which are clearly documented in the product information and quantified

erroneous results which are clearly documented and quantified in the product information and in

7

documented in the product information and quantified in the technical documentation and are

quantified in the product information and in the technical documentation and are

8

to trend reporting pursuant to Article 88; (b) any field safety corrective action

to trend reporting pursuant to Article 83; (b) any field safety corrective action

9

electronic system referred to in Article 92. 2. As a general rule, the

electronic system referred to in Article 87. 2. As a general rule, the

10

as referred to in point (a) of paragraph 1 immediately after they have established the

as referred to in point (a) immediately after they have established a

11

1 immediately after they have established the causal relationship between that incident and

(a) immediately after they have established a causal relationship between that incident and

12

that such causal relationship is reasonably possible and not later than 15 days

that such causal relationship is reasonably possible, and not later than 15 days

13 more passages not shown; read both on EUR-Lex.

Aligned sub-provisions

Medical Devices Regulation first, then IVDR, each with its word ratio.

Changes since the texts were paired

The record holds no change to either provision after the versions they were paired on.